The current chemistry information sheet (cis) pn 988626 af august 2010 for immage complement c4 states that customers may store samples frozen but does not specify a maximum frozen sample storage time. based on the internal testing conducted bec is updating the cis to include a maximum frozen storage time of no longer than 7 days.