Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Manufacturer has confirmed the potential for over-recovery of the c-peptide cvm top level with cvm lots 012 (level j) and 0101 (level 4) when used with the immulite systems c-peptide assays. in some instances this may lead to results outside the range published in the cvm instructions for use (ifu). internal testing demonstrated that all other levels for cvm lots 012 and 0101 are unaffected. the potential over-recovery is limited to the top level cvm and patient results are not impacted.