원인
Manufacturer has determined that it takes 24 minutes of incubation prior to process the samples on the immulite/immulite 1000 system for the patient sample and the pretreatment solution to reach full equilibration when using pretreatment solution (lgfa) lot 055 contained in igf-i kit lot 411. if patient samples are run before reaching full equilibration an under-recovery of up to -36% may occur. if the sample is left to incubate for more than 24 minutes prior to testing no under recovery is observed.