Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Siemens healthcare diagnostics received customer complaints regarding instabilities of high sensitivity c-reactive protein adjusters and controls giving an average positive bias of 20% compared with results from previous lots.