Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Siemens has identified an accelerated decline in counts per second (cps) for specific kit lots of the immulite and immulite 1000 systems third generation tsh assay (catalog numbers lkts1 and lkts5). this may lead to adjustment slopes greater than 1.8 before the expiration date indicated on the kit label is reached. if adjustment slopes are less than 1.8 and qc results are within the established ranges reported patient results are acceptable.