INFANT RADIANT WARMERS 의 리콜

Health Canada (via FOI)에 따르면, 해당 리콜 는 Canada 에서 FISHER AND PAYKEL HEALTHCARE LTD. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    93986
  • 사례 위험등급
    III
  • 사례 시작날짜
    2009-10-21
  • 사례 국가
  • 사례 출처
    HC
  • 비고 / 경고
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • 데이터 추가 비고
  • 원인
    Notifying users in canada of an updated test procedure for the power-fail alarm of the iw900 series infant radiant warmer. the service manual already includes a requirement for an annual check of the power-fail alarm. the test procedure has been revised to check that the power-fail alarm operates for at least 12 minutes. the revised test procedure accounts for any unexpected degradation over time of the supercapacitor that operates the power-fail alarm.

Device

  • 모델명 / 제조번호(시리얼번호)
    Model Catalog: IW910 (Lot serial: All Infant Warmers in Canada.); Model Catalog: IW920 (Lot serial: All Infant Warmers in Canada.); Model Catalog: IW930 (Lot serial: All Infant Warmers in Canada.); Model Catalog: IW932 (Lot serial: All Infant Warmers in Canada.); Model Catalog: IW934 (Lot serial: All Infant Warmers in Canada.); Model Catalog: IW950 (Lot serial: All Infant Warmers in Canada.); Model Catalog: IW980 (Lot serial: All Infant Warmers in Canada.); Model Catalog: ABOVE SERIES WITH SUFFIX JEU (Lot serial: All Infant Warmers in Canada.)
  • 제품 설명
    IW900 Series Infant Warmer