Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Philips has become aware of the following two (2) potential issues that may result in the subframe of the table top becoming free floating and causing unintended horizontal motion. the two potential issues are: 1) service latch may not be secured. 2) service latch threaded rod may be fractured.