Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Some scans unexpectedly set the axial (z axis) scan length to a greater length than the user set in the plan. the issue occurs only when the following 3 conditions are met: -the plan employs dual-surview -the plan requests either a sagittal or a coronal reconstruction -the gantry is tilted.