데이터 추가 비고
원인
When using the roche cobas e411 (rchce41i) driver version 8.00.0001 or 8.00.0002 if there is a barcode read error the current sample may be identified as the last sample id that was in that rack or disk position potentially causing sample mismatch. if the last sample had any pending tests in the orders database test selection would be downloaded to the instrument with the incorrect sample id and results would be reported on the wrong sample.
Model Catalog: 8.00.0002 (Lot serial: Version 8.00.0002); Model Catalog: 8.00.0001 (Lot serial: Version 8.00.0002); Model Catalog: 8.00.0002 (Lot serial: Version 8.00.0001); Model Catalog: 8.00.0001 (Lot serial: Version 8.00.0001)