원인
During beam delivery the tolerance values are used to do a check for unwanted movements. these values are known as static tolerances which are related to all the digital accelerator geometric movement axes except table column. the problem is that the static tolerances from the calibration files (loaded database) are used for beam delivery instead of the machine calculated values. a fault occurs if the loaded database is incorrect.