원인
A software defect that could potentially result in a drop in oxygenation (hypoxia) that is not immediately detected by the clinician. defect involves the use of the "change bed label" feature at the information center ix. if the configuration for this item is set to "optional" and a user selects the "location" button within the "manage patient" menu and makes a change to the patient location alarms for sp02 and/or non invasive blood pressure (nbp) will become disabled without visual notification (bell with an "x") to the user. this problem only happens when the patient is monitored using a networked intellivue trx telemetry m4841/trx4851a patient worn device (pwd).