Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Philips has become aware that the cross reference lines of transversal images from spinal scans are displayed in an incorrect position on a fused image which has been generated through mobiview post processing. the cross reference lines are correctly displayed on the unfused stations. the problem can occur when the field of view of the clinical scan stations is below (more towards the feet) the light visor position. in such situation an error is made in calculating the position information. this causes incorrect positioning of cross-reference lines on fused sagittal images.