IPS CASEDESIGNER 의 리콜

Health Canada (via FOI)에 따르면, 해당 리콜 는 Canada 에서 NOBEL BIOCARE CANADA INC 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    96899
  • 사례 위험등급
    II
  • 사례 시작날짜
    2017-12-11
  • 사례 국가
  • 사례 출처
    HC
  • 비고 / 경고
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • 데이터 추가 비고
  • 원인
    A bug has been discovered via a support case for the ips casedesigner software on november 24 2017 which might lead to designing an incorrect intermediate splint and the user to place the maxilla (and as a result of the occlusion matching also the mandibula) incorrectly during surgery in a very specific scenario: 1.A surgery is planned in the version 1.2 of ips casedesigner using the maxilla rotation point function - moving it from its default position. 2. the user updates the software to the version 1.3. 3. the user opens the planning previously made in the version 1.2 with the version 1.3: the maxilla rotation point will move back to its default position changing the plan originally made by the user. 4. the user saves the planning in the version 1.3. 5. the user orders the splints (intermediate and final) according to the revised planning without noticing the change. the modification of the surgery planning by the sfotware due to the buy can only be detected if the change to the original planning is obvious. the user should always carefully check the surgery planning before proceeding to creating the surgical splint file. a total 37 articles were deployed to 20 customers. only one article 300011_v1.3_win64 was sold in canada.

Device

  • 모델명 / 제조번호(시리얼번호)
    Model Catalog: 300011_V1.3_MAC (Lot serial: K0000022); Model Catalog: 300011_V1.3_WIN64 (Lot serial: K0000022)
  • 제품 설명
    IPS CaseDesigner