Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
The fuse holder on the transformers assembled on the bottom of a number of olympus wm series mobile workstations (wm-np1 wm-wp1 wm-dp1 wm-sc) can overheat and arc if the fuse carrier is not sufficiently tightened.