Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Philips has recently discovered that under specific 802.11 network conditions a partial re-association of the mx40 wlan monitor to a compatible central monitoring system ("information center") is possible. in this state although the information center provides a visible and audible "no data tele" inop alert the mx40 wlan itself enters telemetry mode i.E. its screen turns off in one minute and local alarming is disabled.