Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Philips has become aware that intellivue mx40 system (part numbers 865350 865351 865352) with software revisions b.00 or b.01 may not resume alarming when the alarms are paused and the pause timer ends. this issue will only occur when teh mx40 device is in monitor mode connected to a philips intellivue information center ix (piic ix) that is not in a specific default configuration.