원인
The directview cr version 5.4 software used with the 24 x 30 cm type gp/pq or hr cassettes creates an image size which is 16 % smaller than true size. as a result images smaller than true size will be printed or displayed to the cr user interface or pacs workstation. this will only occur when the true size option is selected within the image editing menu. other uses are not impacted.