Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
The on/off functionality in the lcsu 4 is controlled by electronic components. one of these components might be damaged. the damage occurs over time when the unit is connected to a power source (battery or external power source) but is not related to frequency of use. a damaged component will result in failure of the on/off functionality: the lcsu 4 cannot be turned off and it may turn on by itself.
Model Catalog: 88006101 (Lot serial: 150505A0836 TO 160901A0226); Model Catalog: 88005101 (Lot serial: 150505A0836 TO 160901A0226); Model Catalog: 88005201 (Lot serial: 150505A0836 TO 160901A0226); Model Catalog: 88006201 (Lot serial: 150505A0836 TO 160901A0226)