The lifepak 15 device may no longer be able to deliver defibrillation energy due to a potential failure of an internal component (relay) installed on the therapy printed circuit board assembly (pcba). a malfunction of the relay can be identified during the user test of the device as described in the chapter titled "maintaining the equipment" within the lifepak 15 operating instructions.