Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
It was found during microport's investigation that the biolox forte 36mm alumina ceramic head was confirmed to measure approximately 2mm shorter than other microport 36mm femoral heads (cobalt chrome and biolox delta ceramic femoral heads). the ultimate hazard would be joint laxity and/or dislocation which could result in revision surgery.