Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
The affected products are currently provided a label that indicates that the final torque applied to the abutment to prevent screw loosening should be 30 n-cm. to prevent abutment screw fractures the abutment torque should be reduced to 20 n-cm.