Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
In rare cases the holder of the pc monitor can break and this will cause an unstable monitor. since the monitor weight is approximately 8kg it is assumed that the monitor drops down on the table or on the floor this can result in a serious injury of the hand or the foot or the operator. only a specific range of systems are affected by this recall. the serial number located in the letter is for a specific system componenet not the mammomat inspiration system in which it is located.