Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Due to an assembly error the (protective) adhesive tabs were not removed during the application of the panel digital control to the head assembly of the units. this may result in insufficient conductivity of the panel's emi ground layer to the chassis/system ground potentially causing performance anomalies from electromagnetic interference. prior to device assembly these conductive pads are covered by protective white tabs to preserve the integrity of the conductive epoxy. these white tabs should have been removed prior to installation of the panel to the head assembly.