The manufacturer have received customer reports of arcing/sparking on the defibrillation electrode leadwire. covidien was also aware that the wire/connector subassembly supplier experienced equipment damage due to misalignment from installed tooling. this could r4sult in arcing sparking or thermal damage to the leadwire which could render the device incapable of delivering the appropriate energy or shock to the patient.
“If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.