원인
Correct improper gantry movement and prevent intermittent backup capacitor failures by changing function controller and correct interlock #64 by replacing the fc3 prom with version 3.0.7.
Model Catalog: 9822685 (Lot serial: TOO MANY TO LIST); Model Catalog: 1940753 (Lot serial: TOO MANY TO LIST); Model Catalog: 9401654 (Lot serial: TOO MANY TO LIST)