One lambda inc. part of thermofisher scientific is conducting a correction of the micro ssp allele specific hla class i dna typing tray-a*02. the primer sets in well 3b and 5c are incorrect which may cause a mistype. any a*02:01 a*68:02 sample will generate a positive band in well 3b and a negative band in well 5c. as a result the hla fusion software will display an a*02:75 a*68:01/23/59n typing.