Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Ocd has received complaints of an incomplete foil seal for specific lots of id-micro typing system gel cards. our investigation confirmed that some mts gel cards exhibited a weaker or a loose foil seal resulting in the potential for drying of the first microtube. using gel cards that exhibit drying in microtube(s) may cause false positive reactions leading to erroneous results.