Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
It has recently been discovered that the storage temperature range for the combi m sensor 82 is stated incorrectly in the oprating manual for the microgas 7650. the operation manual incorrectly states that the storage temperature range is -40c to +70c. the correct temperature range is -10c to +50c.