데이터 추가 비고
원인
Beckman coulter has confirmed falsely negative results for voges-proskauer (vp) (pale-pink/brown/colorless) with quality control (qc) american type culture collection (atcc) organism staphylococcus aureus atcc 29213 in the affected panels. the expected result is positive. the falsely negative result was traced to a single vp componenet lot used in the manufacturing of the affected panels.