원인
Medtronic powered surgical solutions testing of the sagittal saw rocker arm seal identified a potential for leakage of fluid into the motor during use. there have been no reports from the field of leakage into or out of the sagittal saw. in assessing the risk of this potential failure mode it was determined that the gravity sterilization cycle included in the ipc user manual may not be adequate to ensure sterility of the saws if fluid ingress occurs beyond the rocker housing seal.
“If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.