원인
When creating 3d plans using dose weighting mode if the user rescales the plan selects "reset" and then changes the wedge angle the plan dose and monitor units (mu) should be returned to the original values before the rescale. instead the system correctly restores the edited wedge field but incorrectly applies the scale reset value again to all other beams in the plan.