Failure to retract or park the varian mlc leaves prior to use of the demountable brainlab m3 could result in the varian mlc leaves interfering with treatment field defined by brainlab m3 without an interlock being asserted. when a demountable brainlab m3 mlc is installed on a varian accelerator the use of two a-b switches is required to deploy the desired mlc. both switches must be set to the correct position prior to use. the first a-b switch box connects the mlc workstation to either the varian mlc controller or to the brainlab m3 controller. the second a-b switch box connects the clinac console to either the varian mlc controller or the brainlab m3 controller. this connection includes the mlc interlock used to inhibit radiation from the clinac accelerator. the mlc that is not selected on this a-b switch is disconnected from the interlock circuitry.