원인
The model 8840 physician programmer may display an incorrect "schedule to replace the pump by" date for the model 8637 synchromed ii implantable drug infusion pump. in some circumstances after a pump's eri has occurred the date may be displayed as a series of question marks or as a date greater than 90 days from the eri date potentially leading to the pump reaching end of service prior to replacement. this issue may only occur if the 90days after eri notification falls on the first day of a month. it does not affect pump function or alarms. this is due to a software error specific to the model 8870 application card software for the synchromed ii pump.
“If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.