원인
Stryker instruments has received reports of casters failing in the field. the caster failures have resulted from the "swivel head of the caster which is a double ball bearing composed of polyamide that loosens over time" creating excessive swivel play. the resultant swivel play allows the caster to bind/jam with the chassis of the rover when force is applied by the operator to move the device.