Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Olympus canada inc. is initiating this recall following investigation of customer complaints associated with water leaking from the irrigation tubing used on the ofp pump. olympus requests that customers inspect the tubing installed on the ofp pump for any signs of water leakage. if they discover any signs of water leakage from the tubing customers should discontinue use of the tubing and replace it with new irrigation tube. olympus has redesigned the irrigation tubing to prevent water leakage.