원인
Onetouch ping insulin pump does not recognize the leap year date of february 29. if the time and/or date were changed on february 29 the time and/or date would not be saved and the pump will revert to the previously saved time/date.
Model Catalog: 10155051 (Lot serial: ALL LOTS); Model Catalog: 10155151 (Lot serial: ALL LOTS); Model Catalog: 10155251 (Lot serial: ALL LOTS); Model Catalog: 10155451 (Lot serial: ALL LOTS); Model Catalog: 10155551 (Lot serial: ALL LOTS)