원인
Opti medical investigated a report from the field where an opti cca-ts2 was found to measure thb high in whole blood. internal investigation has reproduced this finding showing a positive bias in thb results in whole blood when saturated oxygen (so2) is low. this issue was found on all opti blood gas analyzers manufactured after july 2009.