원인
The optical guidance platform may not be properly enforcing a 24 hour time limit between optical camera recalibrations. the optical guidance camera has not been evaluated for stability beyond a 24 hour time period and must be recalibrated at least once every 24 hours. this issue is most likely to be found in units upgraded to the 2.6.X software as older versions of the software allowed users to set calibration time outs greater than 24 hours. however all users should check their settings as described. in particular this issue may be found in sonarray systems where users may have increased the calibration time out beyond the factory settings.