원인
Optima mr360 1.5t: in the event of a significant power surge such as one that might be caused by a lightning strike or a utility equipment failure there is a possibility that the fuse in the mru could be affected. this situation could indirectly lead to a potential hazard if an emergency magnet quench is required.