Upon unplugging the centrifuge from the wall socket there is a possibility for a hazardous voltage (>42.4 v peak) to remain present at the prongs of the instrument power plug beyond 5 seconds. when removing the device's power plug from the power source there is a possible risk of electric shock if the user were to touch the prongs of the power plug.