Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
On october 6 southmedic received a customer complaint in which the co2 diffuser of the etco2 mask became disconnected from the mask andfell into the patient's mouth. further inspection of the identified lot number and investigation into root cause identified the potential for loose or missing diffusers in other lots.