원인
Somagen diagnostics along with thermofisher scientific would like to notify you of an urgent recall and backorder of immunocap allergen c5 ampicilloyl product code 14-4450-01 and immunocap allergen c6 amoxicilloyl product code 14-4451-01. an internal monitoring stability study has indicated that stability specifications have not been met and were in fact half the intended expiry date of 12 months. there is a possibility that samples with a low positive concentration may be incorrectly reported as negative. any reported test results below the assay limit of qualification (loq) should be disregarded as these results may be invalid. no complaints or injuries to patients have been reported.