원인
The door roller assembly on the plum lifecare 5000 series and plum xl infuser families has the potential to break which can lead to possible unrestricted flow and/or overdelivery when the door of the pump is opened.
Model Catalog: 11859 (Lot serial: All Serial Numbers); Model Catalog: 11846 (Lot serial: All Serial Numbers); Model Catalog: 11781 (Lot serial: All Serial Numbers); Model Catalog: 11555 (Lot serial: All Serial Numbers); Model Catalog: 11845 (Lot serial: All Serial Numbers); Model Catalog: 2507 (Lot serial: All Serial Numbers)