원인
During internal stability testing of the reconstituted precicontrol multimaker of lot 178364 a reduced stability of pigf became apparent. whereas the stability claim for all analytes in precicontrol multimaker is 72 hours at 2-8c this claim does not apply for pigf. as a consequence the stability of pigf has been reduced accordingly in the method sheet of the precicontrol multimaker. the claim at 2-8c was removed but the frozen and ambiant temperature stability remained the same as they are not affected.