Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
The manufacturer (microlife corp.) incorrectly installed a component on the printed circuit boards in (2) lots of the probp 2400 during rework in the manufacturing process. as a result of the defective electrical component in the battery charging circuit when the probp 2400 is connected to the external power supply/battery charger a potential over-voltage battery charging condition may occur. this potential defect resides with the device not the battery. the over-voltage condition can result in high nimh battery temperatures that in some cases can reach levels sufficient to cause melting of the plastic (abs) battery door and other adjacent plastic and foam device components.