데이터 추가 비고
원인
This field action is being initiated to correct a recently identified real time controller (rtc)firmware error found in firmware versions 3.0 3.1 3.2 and 3.3. this error may cause the column of the vantage panoramic x-ray unit to continue moving vertically despite the release of the movement control button (i.E. "failure to stop"). if this occurs while a patient is positioned under the column there is the potential for injury such as bruising sprains or resulting from quick movements to avoid the moving column.