원인
Philips has become aware that if the wifi connection between the skyplate detector and hp transfer point is weak an image may fail to transfer from the skyplate detector to the system. the image remains in the memory of the detector but cannot be transferred wirelessly or by use of the backup cable. to continue the operator can reset the skyplate detector by removing its batteries but the acquired image is lost and a re-take is necessary. (see e-file for french translation).