Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
A limited number of protime instruments were manufactured with a component that may contribute to a small decrease in performance and may cause inconvenience. affected instruments have a potential to yield either an error code (no test) or a slightly lower than actual pt/inr result when the inr is greater than 3.5.