Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
The barcode on the quality control certificate that accompanies the product provides the incorrect control name when using the barcode scanner to add a new lot of controls. the barcode on the certificate does not match the control tube labels. the quality control tube labels display the correct description "l" and "h" for the low control and high control respectively while the quality control certificate indicates the incorrect control names "dsdnan" for the negative control and "dsdnap" for the positive control. users are still able to use the controls but will need to manually enter the correct control name.