Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Potential issue with qube compact monitors that are "docking station ready" - that have had a port cover plate removed allowing the monitor to be used with the optional qube docking station. if a docking station ready qube monitor is powered up undocked connected to the network via an ethernet hardwire connection and then disconnected from the ethernet hardwire connection and inserted into a qube docking station a switch condition will cause the monitor to lose the network connection. all bedside alarms will continue to function normally. no one has been injured as a result of this issue.